General Specifications:
Modular system: Quick and easy approach for upgrades, repairs and reconfiguration of the whole system
Rapid pipetting speed minimises assay drift, ensuring consistent results across the plate and plate-to-plate
Learned error recovery: Supporting walk-away automation with trained error recovery actions
Automatic cover lock protecting reagents and samples from light and interferences
LIS and LIMS compatible
QC monitoring
Sample identification with an optional on-board barcode reader
Disposable tips eliminating carryovers
Compatible with DIALAB® ELISA assays
No. of plates: 4
No. of assays: 1 assay per strip or up to 12 assays/plate
No. of samples: 96
No. of reagents: 24
No. of Std/Ctrl: 33
No. of pipettes: 1
System modules
Reader
Washer
Incubators (Up to four)
Sample module
Ambient drawer
Barcode reader (Optional)
Liquid tanks for system - and waste water
Sample Pipetting
Sample tip size: 300 μL
Sample pipetting volume: 5 – 250 μL; 5 – 250 μL single-shot; 25 – 100 μL multi-shot
Precision single-shot: ≤ 3% CV at operating volume above 10 μL
Accuracy single-shot: ± 2% of target volume at any operating volume above 10 μL
Dilution range: 1 part in 190 one-stage dilution; 1 part in 36100 two-stage dilution
Number of sample-tips: 4 racks of 108 tips
Sample tube dimensions: 12 – 16 mm Ø external dimension; 55 – 100 mm length
Reagent Pipetting
Reagent tip size: 1300 μL
Number of reagent tips: 41
Reagent pipetting volume: 35 – 1000 μL
Reagent pipetting precision: ≤ 3% CV at 10 shots at any volume in operating range above 25 μL
Reagent pipetting accuracy: ± 2% of target volume in operating range (singe-shot mode)
Washer Specifications
Manifold configuration: 8 – way
Programmable volumes: 50 – 999 μL
Wash containers: 4 wash bottles at 2.0 L with level-sensing
Waste containers: 8.0 L with waste-full sensor
Residual wash volume: < 3 μL per well with dual-axis sweep in a flat-bottom plate
Dispense precision: ≤ 5% CV (with 300 μL in a 96 well plate)
Incubator Specifications
Number of incubators: Up to 4
Temperature range: RT, 7° C to 50° C
Temperature accuracy: ± 1° C
Shaking: > 15 Hz periodic or continuous
Reader Specifications
Photometric range: 0.000 – 3.000 OD
Spectral range: 405 nm – 690 nm
Filter: 405 nm, 450 nm, 490 nm, 620 nm
Precision: ± 0.01 OD at 0.000 – 0.500 OD; < 1% CV at 0.501 – 2.000 OD; < 1.5% CV at 2.001 – 2.500 OD
Accuracy: ± 0.01 OD or 2.5% (0.000 – 3.000 OD)
Reading time: < 10 seconds, single wavelength; < 20 seconds, dual wavelength