General Features
Clotting and immunoturbidimetric assays
Throughput: PT 137 tests/hour, APTT 60 tests/hour ± 20 tests/hour
Micro method
STAT program
QC program
Bidirectional LIS
Continuous loading of samples
Internal sample barcode reader
Color touch display 17.0 x 12.5 cm TFT
ChipCARDTM for cuvette management
Technical Specifications
Calibration points: Up to 7 calibration points, 2 calibration curves per test or automatic calibration
LIS compatible: Yes
Measuring principle: Photometric, 2-channel LED photometer with 2 reference channels, Wavelengths: 450 nm, 570 nm, 740 nm
Temp. controlled measuring unit: 37.0°C ± 0.4°C
Sample: Plasma
Result conversion: Seconds, INR, %, Ratio, Units (mg/dL, g/L)
Cuvettes: Disposable cuvette ring with 32 cuvettes
Sample positions: 2x 11 positions
STAT positions: 3 positions
Reagent positions: 15 positions (3x 15 mL, 12x 5 mL), 3 of them stirred
Cal & Ctrl positions: 8 positions
Pipettor washing station: Yes
Sampling: Capacitive pipettor heated at 37 °C
Dilutor: 2 – 250 µL (precise in 1 µL steps)
Language: English, German, Russian, Spanish
Printer: External
Interface: 2x RS232C, 2x USB, LAN & ChipCARD™ reader
Methods: PT, Factors (II-XII), Factors V Leiden, D-Dimer, Protein C&S, Ecarin, FIB, PT/FIB, BT, TDT, APTT, PiCT, TT, AT III. Up to 50 methods programmable